GEF-H1, Proteintech, 24472-1-AP
Information about this antibody
Vendor notes
User Reviews (1)
Review of: GEF-H1, Proteintech, 24472-1-AP by pAbmAbs Data Mining
Our team has carried out deep research in the published scientific literature to identify papers and reports using antibodies directed against human GEF-H1 and followed up by extraction and evaluation of the available data to assess its performance according to pAbmAbs rating criteria.
YCharOS is here conducting a large head-to-head study comparing 11 commercial research antibodies against human GEF-H1 encoded by ARHGEF2.
The collection of antibodies was tested using U-87 MG WT and ARHGEF2 KD cells in the applications WB (WT and KD cell lysates), IP (WT cell lysate) and IF (WT and KO cells) according to YCharOS standard protocols (see reference).
In WB (1:1000 ab dilution) 24472-1-AP detected a band specific to WT cell lysate with no corresponding band observed in KD cell lysate. Furthermore, faint bands of lower molecular weight were observed in both genotypes. A band was detected following IP (2 µg Ab used), this band was markedly more intense compared to band in input extract (6% of starting material) and no band was detected in immunodepleted extract (6% of total unbound fraction), showing that the antibody does completely deplete the antigen from cell lysate.
In IF (1:1000 ab dilution), staining was observed in WT cells only.
Comments