Glucagon receptor, Thermo Fisher Scientific, PA5-77446
Information about this antibody
Vendor notes
User Reviews (1)
Review of: User contributed: Glucagon receptor, Thermo Fisher Scientific, PA5-77446 by pAbmAbs Data Mining
We here review a publication from Novo Nordisk and University of Copenhagen providing a head-to-head comparison of all commercially available antibodies against the glucagon receptor (GCGR) (Bomholt et al., Communications Biology, 2022).
The authors systematically evaluated 12 commercial antibodies (Table S1) in HEK293 cells transfected with mouse or human cMyc-tagged GCGR cDNAs and studied liver sections from Gcgr+/+ and Gcgr-/- mice. Autoradiography, RNA-sequencing, and single cell RNA-sequencing were used as antibody independent approaches to support the findings obtained with IHC.
Transfected cells were permeabilized with methanol.
All GCGR antibodies, including this one from Thermo Fisher, gave overlapping staining with the cMyc antibody, demonstrating the ability to specifically recognize recombinant GCGR in transfected cells.
No staining was seen in control transfected cells or cells without primary antibodies (Fig. 1).
The antibodies were also checked for their ability to detect membrane bound GCGR by staining unpermeabilized HEK293 cells transfected with human GCGR and fixed in 4% PFA.
Thermo Fisher PA5-77446 (Antibody 9 in the paper) also showed positive staining of membrane bound GCGR (Fig. S1)
The antibodies were subsequently evaluated using formalin-fixed and paraffin-embedded liver sections from female Gcgr+/+ and Gcgr-/- mice. However, the Thermo Fisher antibody showed no staining (Fig. 2a)
The antibody also showed no signal in IHC on paraffin embedded human liver tissue from patients with NASH (Fig. 2S).
In conclusion, Thermo Fisher PA5-77446 receives a 3 star rating in immunostaining based on its ability to specifically recognize GCGR in transfected cells.
We highly encourage the scientific community to share their experiences.