User Reviews (1)
Review of: Rab6A, Abnova, H00005870-M01 by pAbmAbs Data Mining
Our team performed a comprehensive review of the scientific literature to identify studies using antibodies against human Rab6A, followed by extraction and evaluation of available data according to pAbmAbs rating criteria.
YCharOS conducted a systematic head-to-head comparison of 13 commercially available antibodies against Rab6A across multiple applications
Antibodies were tested by WB using lysates (40 µg protein) from HAP1 WT cells and HAP1 Rab6A KO cells, which may express a truncated lower molecular weight variant. These cell lines were also used for IF. In addition, antibodies were evaluated by IP using HAP1 cell lysates (1 mg protein). All experiments were performed according to YCharOS standard protocols (see reference).
In WB (1:1000 antibody dilution), H00005870-M01 detects a weak band at the expected size (~24 kDa) in WT HAP1 cell lysate with no corresponding band in lysate from Rab6A KO cells. In addition several weak background bands are detected in both lysates.
In IP, the antibody (2 µg) does not enrich Rab6A following precipitation, as detected by WB using the Rab6A-specific antibody NBP3-13701. Rab6A in the starting material, unbound fraction, and precipitate was assessed. An intense IgG LC signal is observed in the IP fraction
In IF, the antibody (1:500 dilution) resulted in a weak staining of HAP1 WT as well as Rab6A KO cells.
We highly encourage other scientists to share their observations.
Comments