Review of: SLC25A11, Thermo Fisher Scientific, PA5-96958 by pAbmAbs Data Mining
Our team has carried out deep research in the published scientific literature to identify papers and reports using antibodies directed against human mitochondrial 2-oxoglutarate/malate carrier protein and followed up by extraction and evaluation of the available data to assess its performance according to pAbmAbs rating criteria.
YCharOS is here conducting a large head-to-head study comparing 6 commercial research antibodies against human mitochondrial 2-oxoglutarate/malate carrier protein encoded by SLC25A11.
The collection of antibodies was tested using HEK293T WT and SLC25A11 KO cells in the applications WB (WT and KO cell lysates/25 µg protein) and IP (WT cell lysate) according to YCharOS standard protocols (see reference).
In WB (1:200 ab dilution) PA5-96958 detected no band specific to WT cell lysate. No band was detected following IP (2 µg Ab used).