Our team has carried out deep research in the published scientific literature to identify papers and reports using antibodies directed against human Calponin-3 and followed up by extraction and evaluation of the available data to assess its performance according to pAbmAbs rating criteria.
YCharOS is here conducting a large head-to-head study comparing 8 commercial research antibodies against human Calponin-3 encoded by the CNN3 gene.
The collection of antibodies was tested using HAP1 WT and CNN3 KO cells in the applications WB (WT and KO cell lysates/20 µg protein), IP (WT cell lysate) and IF (WT and KO cells) according to YCharOS standard protocols (see reference).
In WB (1:1000 Ab dilution) MA5-38502 detected a band of appr. 36 kDa in WT cell lysate, with no corresponding band in KO lysate. No band was detected following IP (2 µg Ab used). In IF (1:1000 Ab dilution), comparable staining intensities were observed in both KO and WT cells.
We highly encourage other scientists to share their observations.