Review of: ADNP, Thermo Fisher Scientific, 702911 by pAbmAbs Data Mining
Our team has carried out deep research in the published scientific literature to identify papers and reports using antibodies directed against human ADNP and followed up by extraction and evaluation of the available data to assess its performance according to pAbmAbs rating criteria.
YCharOS is here conducting a large head-to-head study comparing 7 commercial research antibodies against human ADNP.
The collection of antibodies was tested using DMS 53 WT and ADNP KO cells in the applications WB (WT and KO cell lysates/30 µg protein), IP (WT cell lysate) and IF (WT and KO cells) according to YCharOS standard protocols (see reference).
In WB (1:1000 Ab dilution) 702911 detected a band of appr. 135 kDa in WT lysate, with no corresponding band observed in KO lysate. Only a very weak band was detected following IP (2 µg Ab used), together with two bands of lower molecular weight. In IF (1:500 Ab dilution), comparable staining intensities were observed in both KO and WT cells. We highly encourage other scientists to share their observations.