Ataxin-2, Thermo Fisher Scientific, PA5-53775
Information about this antibody
Vendor notes
User Reviews (1)
Review of: Ataxin-2, Thermo Fisher Scientific, PA5-53775 by pAbmAbs Data Mining
Our team has carried out deep research in the published scientific literature to identify papers and reports using antibodies directed against human Ataxin-2 and followed up by extraction and evaluation of the available data to assess its performance according to pAbmAbs rating criteria.
YCharOS is here conducting a large head-to-head study comparing five commercial research antibodies against human Ataxin-2 encoded by ATXN2.
The collection of antibodies was tested using HAP1 WT and ATXN2 KO cells in the applications WB (WT and KO cell lysates/50 µg protein), IP (WT cell lysate) and IF (WT and KO cells) according to YCharOS standard protocols (see reference).
In WB (1:1000 Ab dilution) PA5-53775 detected a band of appr. 140 kDa in both WT and KO lysates. Band in KO had reduced intensity. Several background bands of both high and lower molecular weight were detected in lysates of both genotypes. A band was also detected following IP (1 µg Ab used). This band was more intense than the corresponding band in input extract (10% of starting material). Band was also detected in immunodepleted extract (10% of total unbound fraction), showing that the antibody does not completely deplete the antigen from the cell lysate. In IF (1:400 Ab dilution), KO showed less staining than WT, but background signals was still detected in KO.
We highly encourage other scientists to share their observations.